Back to Product Guide - Frequency/Rate/Total/Timer/Batch
6-DIGIT FREQUENCY METER, TACHOMETER, RATE METER, TIMER, PULSE TOTALIZER, PROCESS METER & TOTALIZER WITH RS-232
PENTA P6000

Made in the USA
Electromagnetic Compatibility
Year 2000 Compliant

Price List
Mechanical Reference
1/8 DIN Case
1/16 DIN Case
Mechanical Reference
Connections Diagram
OPERATING MODES
1. Frequency/tachometer from 10-6 Hz to 7 MHz
2. Frequency ratio FB/FA from 0.02 Hz to 7 MHz
3. Totalizer up or down from ­99,999 to 999,999 or accumulating stopwatch
4. Period/period average from 140 ns to 36 days
5. Time interval/time interval average/stopwatch
6. Process meter for 4-20 mA, 0-2 or 0-10 V input
7. Integrating totalizer for 4-20 mA, 0-2 V or 0-10 V input

FAST LOW-FREQUENCY MEASUREMENT
Every 50 ms + 1 signal period
Slower read rates for frequency averaging

PROGRAMMING FEATURES
Scale factor from ­99,999 to 999,999 (any decimal point, multiply or divide)
Offset from ­99,999 to 999,999 (any decimal point)
HI and LO setpoints for control or alarm
Programming via front-panel or RS-232
Program stored in non-volatile memory
Four levels of program lockout for security
Fixed decimal point or autoranging

SIGNAL CONDITIONER CHOICES
Dual-channel TTL with protection to 25 V
Single-channel, isolated, with excitation
Dual-channel, isolated, with excitation
Single-channel, non-isolated, with excitation
Analog input, isolated, 4-20 mA, 0-2 V, 0-10 V

COMMUNICATIONS & CONTROL
RS-232 or 20 mA serial ASCII output (std)
HI, LO, GO 150 mA open-collectors (std)
Dual 8A Form C relays (opt)
Parallel BCD output, isolated (opt)
Analog output, isolated and scalable, 4-20 mA, 0-20 mA, 0-10 V (opt)

DISPLAY & MECHANICAL
Six 0.56 in (14.2 mm) 7-segment LED characters
Five-key programming front panel (std)
Plain front panel (opt)
Screw-clamp connectors for signal and power
1/8 DIN case

In its base configuration, the P6000 is a microprocessor-based, 6-digit, 1/8 DIN counter which can be configured by front-panel keys or by a personal computer as a frequency meter/tachometer, frequency-ratio meter, period/period- average meter, time-interval/time-interval-average meter or totalizer. It combines these five operating modes with ease of setup, wide dynamic range, six-figure crystal-based accuracy, and software scaling.

With the addition of an optional analog-to-frequency signal conditioner, the P6000 can become a software-scalable process meter with two setpoints and exceptionally wide zero offset capability. It can also become a 6-digit analog integrating totalizer.

The P6000 provides a five-key front panel, which can be used to select mode of operation, scale factor, zero offset and two setpoints for ON/OFF control or alarm. Setup parameters can be saved in non-volatile memory with four levels of front-panel lockout for program security. In addition, the P6000 can be programmed via RS-232. It can also report its own setup data and transmit ongoing readings and alarm status via RS-232 or 20 mA ASCII current loop. Modem support is built in for remote operation.

FLEXIBLE SIGNAL CONDITIONING

0. TTL-LEVEL PULSE INPUTS
Dual non-isolated TTL/5 V CMOS-level input channels with protection to 25 V dc are standard and can accommodate frequencies up to 7 MHz. The inputs can be tied to contact closures by using a spare flip-flop available at the connector for debounce. They can also be tied to sensors with an open-collector NPN or PNP output if these are powered externally. Contact closures require an external 20 kOhm pull-up resis-tor. PNP sensors require an external 1 kOhm pull-down resistor.

1 & 2. ISOLATED SIGNAL CONDITIONER WITH EXCITATION OUTPUT
This almost universal signal conditioner is available in single- or dual-channel versions. It provides sensor excitation output plus AC or DC coupling, signal isolation to 350 Vp, and jumper-selectable low-pass filtering, debounce time and hysteresis. It allows the P6000 to be tied directly to passive magnetic pickups with output down to ±10 mV, to AC line voltages up to 240 V rms, and to NPN, PNP, NAMUR or contact-closure sensors all with a high degree of input protection.

3. NON-ISOLATED SIGNAL CONDITIONER WITH EXCITATION OUTPUT
This is an economical single-channel non-isolated signal conditioner which supplies power up to 16 V at 25 mA for direct 3-wire connection to NPN sensors, or 2-wire connection to NAMUR sensors (<1 mA ON, >3 mA OFF) and contact closures. It can also be used with magnetic pickups and other active voltage sources from 0 to 200 mV up to 60 V rms.

Frequency Measurement
The P6000 measures frequency or period by counting the number of 11.059 MHz clock pulses during an actual gate time t, which corresponds to an intergral number of signal periods N. This technique allows high-accuracy low-frequency measurements. Frequency is calculated from N/t, period from t/N.

4. ISOLATED ANALOG-TO-FREQUENCY SIGNAL CONDITIONER
This signal conditioner accepts 4-20 mA, 0-5 V or 0-10 V analog signals and turns the P6000 into a process meter with isolated input, 6-digit scale and offset capability, two setpoints and RS-232. It also allows the P6000 to serve as an analog integrating totalizer, for instance to display volume based on the 4-20 mA signal from a flowmeter.

More
Fast Quotes on Products
HOME / PRODUCT GUIDE / TECHNICAL REFERENCE